TOMY AUTOCLAVES SX 700 Specifications Model name : SX-700 Operating Temperature Range Sterilizing : 105 < 135 C 0.019 < 0.212MPa Heating : 45 < 104 C 0 < 0.015MPa Warming : 45 < 95 C Maximum operating pressure : 0.25MPa Temperature Display : Digital Pressure Gauge Display : Analog Display range : 0 < 0.4MPa Heat source : 3kW electric heater Safety device : ? Water level sensor, ? GFCI, ? Lid interlock, ? Over-heat prevention, ? Over-pressure prevention, ? Open temperature sensor detection, ? Safety valve Time Display : Digital Display range Sterilizing : 0: 00 - 9: 59, 0 - 99hours, 0 - 999min. Heating : ( 0: 00 - 9: 59, 10 - 99hours changeable) Warming : Variable: 0 - 99hours Chamber dimensions : 370 x 774mm Chamber capacity : Effective internal volume: 69L, Internal volume: 79L Chamber material : SUS304 Dimensions ( mm) : 470W x 528D x 1003Hmm ( With protruding: 625D) Weight : 72kg Rated voltage : - 230V AC Power input : - 13A/ 230V Required power supply : - Single-phase 230V AC ( 50/ 60 Hz) 15 A or more Power consumption ( calorific value) : 3.0kW ( 2580cal/ h) Environmental Conditions : When operating the autoclave, observe the environmental conditions given below. Ambient temperature: 10 to 35 C Atmospheric pressure: 860 to 1060hPa Relative humidity: 30 to 85% ? Maximum gradient: 2 Accessories : Stainless baskets 2 ( 345 x 181mm) ? Chamber bottom Plate 1, ? Caster stoppers 4, ? Operator' s Manual 1, ? Clear folder ( for storing the operation manual) 1, ? Screw ( for attaching the clear folder) 1, ? Warranty card 1, ? Customer card 1, ? Inspection Sheet 1
For More Info :